Diascan RESCHEDULED TWICE
Press SELECT key to confirm
Appears:
Appears when two Diascan measurement checks in a row
has been interrupted by any of the following actions:
-
the blood
fl
ow rate is changed
-
the
Blood pump button
is pressed
-
the
Start UF Stop button
is pressed
-
the
Fluid bypass button
is pressed
-
a treatment alarm which produces dialysis
fl
uid
bypass or blood pump stop is generated
-
sodium and bicarbonate (conductivity) of the dialysis
fl
uid is changed
-
temperature of the dialysis
fl
uid is changed
-
dialysis
fl
uid
fl
ow rate is changed
-
UF volume is changed
-
linear UF pro
fi
ling settings is changed
-
linear UF pro
fi
ling activation/deactivation is made
-
sodium or bicarbonate pro
fi
ling activation/deactiva-
tion is made
-
isolated UF activation is made
The machine was not able to completely perform the
measurement check as the measurement check has been
interrupted.
Machine actions:
None.
14:44
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...