insuf
fi
cient inlet pressure. The function check will wait for water
from the WRO before continuing.
8. Switch the machine off by pressing the
On/Off button
for
3 seconds.
If a new treatment is to be started directly after the disinfection
program, press the
fl
ashing
Priming button
brie
fl
y and
PRESS SELECT 3 SECONDS TO START FCH
will be
displayed. Press the
Select key
for 3 seconds to start a new
function check and simultaneously reset previous values to zero.
Manual start of the Low Flow Heat phase
1. Press the
Rinse/Disinfection button
brie
fl
y.
2. The
RINSE/DISINFECTION MENU
will be displayed on
the Information Display.
3. Select
HEAT
.
4. Check that
HEAT
is displayed as selected program. If
HEAT
is displayed, continue to point 4. If not, do point 4.a
and 4.b before continuing to point 5.
4.a.
Select
CHANGE
.
4.b.
Select
HEAT
.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Hygiene and Maintenance
8:21