
5. If desired, adjust the automatically set venous pressure alarm
limits as necessary using the
Keypad
. Start by adjusting the
high alarm limit. Then, during the arterial phase, gradually
adjust the low venous pressure alarm limit as well. Make
sure that the venous pressure alarm limits remain within
the limits recommended by the clinic for that particular
patient's blood access. See the next following Note for
further information about this setting.
6. If necessary, adjust the level in the venous expansion
chamber. The level should be as low as possible (i.e. without
air passing into the blood lines at the end of the venous
phase), in order to best achieve a proper stroke volume
(
STROKE
).
7. Check that the current stroke volume (
STROKE
) shows a
proper value. See "Glossary of single needle parameters and
key terms" on page 5:2 in part 1 for further information.
8. Press the
fl
ashing
UF Start/Stop button
(lights up) to start
ultra
fi
ltration. Note especially that at all times whenever
necessary during treatment, the
UF Start/Stop button
can
be pressed (
fl
ashing) in order to decrease the UF rate to
minimum UF rate.
Check that the minimum UF rate is set to the correct value.
Press the
Fluid Path button
, select
UF
and check the value in
SET MIN UFR
. Adjust if necessary.
Note
•
The achieved stroke volume is determined by the size and
position of the venous pressure alarm window. The wider the
window, the greater the stroke volume. The higher the position
of the window, the higher the venous
fl
ow. Set the venous
pressure alarm limits in such a way that a proper stroke volume
is being achieved. Bear in mind that as the venous pressure
alarm limits are used to control the arterial and venous phases
the extracorporeal circuit must continuously be monitored
by visual inspection. A decreasing stroke volume and/or a
decreasing mean blood
fl
ow rate may indicate clotting or a
kinked venous blood line.
To optimize the treatment:
Note
→ → →
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Single Needle Treatment
5:9
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...