To start the air detector test:
Fill the venous drip chamber
Appears:
During the beginning of the priming procedure when the
venous drip chamber has not yet been
fi
lled.
Machine actions:
None.
Measures:
Fill up the level in the venous drip chamber (while the
blood pump is running) using the level adjustment knob
(anticlockwise direction). Adjust the level so it is well
above the air detector head. When
fi
lled, the machine
fi
nalizes the function check and the set treatment time
will appear on the Time Display (on condition that bypass
path of the Flow Diagram is green). For information
on the additional activation of the air detector alarm
function, see in the instructions for the type of priming
procedure being used in chapter 4.
Mute time:
-
Buzzer sound:
No buzzer
14:118
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order