CAUTION
To prevent the syringe from loosening from the heparin pump
during treatment the syringe plunger has to be thoroughly secured.
Check by lightly pulling on the plunger. When doing so, it must be
impossible to pull the syringe out of the heparin pump.
CAUTION
Note
•
Make sure that chosen syringe type matches displayed syringe
type
in the
HEPARIN MENU
, if not call authorized service
technician.
Note
Preparations if the heparin pump is to be used
8.a.
Prepare a syringe with the desired amount of heparin
solution.
8.b.
Attach the syringe to the connection at the end of the
thin line coming out just after the blood pump segment
(pointed out in corresponding
fi
gure) on the arterial
blood line.
8.c.
Prime the line with the heparin solution to the point
where the line ends in the arterial blood line. The luer
lock mounting ensures that the syringe is properly
connected to the line, see corresponding
fi
gure.
8.d.
Pull out the piston on the heparin pump to end
position. To change the position of the piston, press
the end of the piston, hold it in and move the holder.
When the piston is released the holder cannot move.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Double Needle Treatment
4:25
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...