Appears:
The venous pressure has passed the set venous pressure
low alarm limit.
Machine actions:
-
The blood pump stops.
-
The venous blood line clamp closes.
-
The ultra
fi
ltration rate is set to zero and the dialysis
fl
uid is bypassed from the dialyzer (when blood is
detected).
When the venous pressure is within the set alarm limits
again, the blood pump automatically starts and the venous
blood line clamp opens.
Possible cause:
-
The venous blood line has separated from the dialyzer.
-
The venous blood line has separated from the venous
needle.
-
The position of the venous needle has changed.
-
Clotting before or in the dialyzer.
13:16
AK 96
®
Operator's Manual - Alarms
HCEN12239 Revision 12.2010
Program version 3.xx