It is possible to
choose one the two following ways to continue:
1.a.
Con
fi
rm treatment end by pressing the
Select key
.
The
TREATMENT TIME EXPIRED
attention alarm
will be cleared and the
RINSE BACK MENU
will be
displayed on the Information Display. Continue by
following the instructions in the next section, “Return
the Blood”, page , point 2.
1.b.
Con
fi
rm treatment end by pressing the
fl
ashing
Time button
. The
TREATMENT TIME EXPIRED
at-
tention alarm will be cleared and the menu will change
(this menu will automatically disappear after 20 sec-
onds or when the
Blood pump button
is being pressed).
Press the
Select key
and the
RINSE BACK MENU
will
be displayed on the Information Display. Continue by
following the instructions in the next section, “Return
the Blood”, page , point 2.
If the treatment is to be continued at this point, press
the
Back key
and increase the treatment time (
SET
)
using the
Keypad
. Increase the UF volume, or set the
min UF rate to zero.
Return The Blood
CAUTION
During rinse-back procedure, when the blood is returned to the
patient, the operator is responsible for monitoring the venous
pressure since the supervision of venous needle dislocation is
deactivated.
The arterial clamp can be open
during rinse-back procedure by
preset. This is to facilitate returning all blood in the extracorporeal
circuit to the patient. Bear in mind that if this preset is active, the
operator is fully responsible for manually clamping the arterial blood
line and checking it for air whenever necessary during rinse back
procedure.
CAUTION
Procedure
As con
fi
rmation of treatment end has previously been con
fi
rmed
(brie
fl
y pressed
Select key
), the
RINSE BACK MENU
is
displayed on the Information Display.
2. When
RINSE BACK
has been selected, press the
Select key
for 3 seconds to activate discontinuing mode.
The menu will change and request the operator to press the
fl
ashing
Blood pump button
when ready for rinse-back.
4:60
AK 96
®
Operator's Manual - Double Needle Treatment
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...