Proportioning of concentrates
The proportioning of concentrate is done through conductivity
control. The concentrates are pumped into the system with one/two
volumetrically supervised pumps. No minimum feeding pressure is
necessary, the
fl
uid is sucked in. Maximum feeding pressure, 150 kPa.
Acetate
Na
+
, 130 to 160 mmol/l, (±3 mmol/l)
Bicarbonate
Na
+
, 130 to 160 mmol/l, (±6 mmol/l)
HCO
3-
, 20 to 40 mmol/l, (±6 mmol/l)
Measuring range
13 to 16 mS/cm
Accuracy
0.2 mS/cm
Alarm limits
±5 % of the calculated conductivity set
value
Ultrafiltration control
Note
•
Besides the ultra
fi
ltration, patient weight change during
treatment is also affected by other factors. These include
factors such as
fl
uid intake, food intake, perspiration, drug
administration, infusion priming and rinse-back volumes
amongst others. In addition, precise pre- and post-treatment
weight is critical for adequate assessment of the ultra
fi
ltration
during the treatment. If these measurements are not accurate
a discrepancy between the ultra
fi
ltration achieved during the
treatment and the changes in body weight will occur.
Note
Volume control
Direct electromagnetic measurement of
dialysis
fl
uid
fl
ow, before and after the
dialyzer.
UF volume
Adjustable 0 to 10.00 l
Accuracy of measured
volume
± 50 ml or ± 50 ml * passed treatment
time (h) or ± 2.5 % whichever is greater.
UF coef
fi
cient
Maximum 85 ml/h/mmHg
UF-rate
0.0 to 4.0 l/h, given by the set values of
UF volume and treatment time.
Time
Remaining treatment time control. 0.05
to 9.59 hour. minute (±1 minute)
Profiling
UF-rate
0.0 to 4.0 l/h
Na
+
, Acetate/Bicarbonate
mode
130 to 160 mmol/l
HCO
3-
, Bicarbonate mode
20 to 40 mmol/l
9:6
AK 96
®
Operator's Manual - Technical Data
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...