CLEANING REQUIRED
hours since last cleaning
###
Appears:
When a cleaning program is needed.
Machine actions:
None.
Measures:
If possible, run a cleaning program, i.e. a heat
disinfection program with CleanCart-A cartridge or a
chemical disinfection program with sodium hypochlorite.
If not, the attention alarm can be cleared by pressing
the
Select key
. If the machine is not disinfected after
treatment, the attention alarm will reappear during the
next treatment.
The period of time between the last performed cleaning
and the appearance of the attention alarm can be preset.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:31
Attention alarms in alphabetical order