Close the blood pump cover by pressing at the point shown
in the corresponding
fi
gure.
Make sure the blood pump segment collars are outside the
pump housing.
It is recommended to use a blood line with a plastic cross-bar
between the segment collars, pointed out as
in the
fi
gure.
The cross-bar holds the segment in place during treatment
and it is also easier to thread the blood pump segment
through the pump.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Double Needle Treatment
4:21