Environmental data
Operation
If condensation occur when moving the equipment between locations
with different temperatures and high relative humidity ( e.g. outdoor
and indoor locations ), the inside of the equipment shall be allowed to
dry before switching on the equipment.
Ambient Temperature range +18 to +35 °C
Relative Humidity range
15 to 85 % RH
Air Pressure range (atm.
pressure)
700 to 1060 hPa
Transportation and storage
During transportation and storage the equipment has to be kept in
its original packing. If transportation or storage time is more than
15 weeks, the environmental data relating to the operation has to be
followed. The maximum ambient temperature for transportation and
storage in 96 % Relative humidity is +40 °C.
Ambient Temperature range -20 to +70 °C
Relative Humidity range
10 to 96 % RH
Air Pressure range (atm.
pressure)
500 to 1060 hPa
Electromagnetic environment
The AK 96 is intended for use in the electromagnetic environment
speci
fi
ed below. The customer or the user of the AK 96 should assure
that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The AK 96 dialysis machine uses RF energy only for
its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment
RF emissions
CISPR 11
Class B
The AK 96 dialysis machine is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used
for domestic purposes
Harmonic emissions
IEC 61000-3-2
Class A (Not applicable
for 115 V version)
Voltage
fl
uctuations
/
fl
icker emissions
IEC 61000-3-3
Complies (Not
applicable for 115 V
version)
9:18
AK 96
®
Operator's Manual - Technical Data
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...