6. When the
TRANSMEMBRANE PRESSURE CONFIRM
MENU
is displayed on the Information Display, press the
Select key
to con
fi
rm the settings i.e. that the current TMP
value and the automatically set TMP alarm limits are set in
accordance with the current UF rate and the UF coef
fi
cient
of the dialyzer being used.
7. If the heparin pump is started, this will be running concurrent
to the blood pump. If
BOLUS
has been set this will be
administered
fi
rst. After this the heparin solution will be
administered at the set
fl
ow rate
FLOW RATE
. If the heparin
solution
fl
ow rate is set to zero, an attention alarm appears
requesting the operator to con
fi
rm that the heparin pump is
not to be used.
8. If the functions for isolated ultra
fi
ltration, pro
fi
ling, Diascan,
blood pressure measurement or single needle are going to
be used, see the corresponding chapter in this manual for
instructions on settings.
CAUTION
The correct setting of the UF rate is important to avoid backfiltration.
Set a minimum UF rate according to the instruction for the specific
dialyzer.
CAUTION
Pressure Alarm Limits Setting
Alarm Limit Centralizing Function
The machine is equipped with a centralizing function
to simplify
the correct setting of the alarm limits for arterial and venous pressure
and for TMP.
The centralizing function means that when the
Arterial
or
Venous pressure buttons
are
fl
ashing, the buttons are
to be pressed. When pressed, the alarm window will
automatically be set around the current value. For TMP, the
TRANSMEMBRANE PRESSURE CONFIRM MENU
appears, where
the operator con
fi
rms the automatically set TMP alarm limits.
The size of the alarm limit windows, and where on the pressure scale
they are placed, vary for the different parameters and can be preset by
an authorized technician. The default values which the machine is
preset with upon delivery, can be read in chapter 9, "Technical Data
cations" in part 1. If the centralizing function is removed, or
if wide alarm limits range are preset, it is the operator's responsibility
to make sure that the alarm limits are properly set around the current
values during treatment.
Arterial/Venous Pressure
Press the corresponding buttons (when
fl
ashing) to centralize the alarm
limits around current values.
If necessary the alarm limits can be manually set/adjusted within the
preset alarm window. Press the corresponding button and the pressure
menu will be shown on the Information Display. Manually adjust the
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Double Needle Treatment
4:57
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...