The procedure continues with the
fi
rst
fl
ush phase,
ACTIVE PHASE:FIRST FLUSH
is displayed.
10. If the attention alarm
HIGH VENOUS PRESS TEST NOT MADE
appears, the
venous pressure has not exceeded 50 mmHg during the
priming procedure, which means that the machine has not
been able to test the venous pressure measurement function.
Establish a stable venous pressure over 50 mmHg until the
attention alarm disappears. If this attention alarm is left
unattended, unnecessary alarms and technical error will be
generated during initiating of treatment. See also “Venous
Patient Connector” in “Venous Blood Line - Attach”
previously in this chapter.
When the
fi
rst
fl
ush is
fi
nished the blood pump automatically
stops.
11. Press the
fl
ashing
Blood pump button
to restart the blood
pump and to continue the procedure.
The procedure continues with the second
fl
ush phase,
ACTIVE PHASE:SECOND FLUSH
is displayed.
12. When the second
fl
ush is
fi
nished the assisted priming
procedure is complete and the menu will change.
13. Check that the venous blood line is correctly placed in the
priming detector. See "Venous Blood Line - Attach", point 5.
Press the
Select key
to continue.
It is now possible to continue priming (
EXTRA PRIMING
)
or if requested perform a recirculation procedure
(
RECIRCULATION
). It is also possible to go directly to patient
connection (
CONNECT PATIENT
).
Select desired alternative and continue to the corresponding
instruction as follows.
If necessary, select a new priming procedure (
NEW PRIMING
)
where priming will be repeated from the start and the set priming
volume value be set to zero.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Double Needle Treatment
4:49
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...