IMPROPER POSITION OF CHEMICAL
DISINFECTANT CONNECTOR
Appears:
When the chemical disinfectant connector is not in the
correct position.
Machine actions:
Buzzer activated immediately. If the attention alarm is
active during function check when the chemical valve
leakage test is performed the function check will stop. If
the attention alarm is active during priming or treatment
the correct dialysis
fl
uid can not be prepared.
Measures:
Check the placement of the yellow disinfectant connector.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
14:78
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order