4.
Venous Line Clamp
Introduce the blood line into the venous clamp (marked with
a blue dot) by lightly pressing the line into the clamp.
Then pull the blood line lightly downwards; the line goes into
place when it is lightly pressed at the same time.
5.
Priming Detector
It is of the utmost importance that the venous blood line is
correctly placed in the priming detector for the machine to be
able to detect blood after treatment start.
Check carefully that the venous blood line has reached
end position in the venous line clamp
as shown in the
corresponding
fi
gure. For further information see point 9
"Priming Detector", in Blood Part Component Details"
starting on page 2:5 in part 1.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Double Needle Treatment
4:33