5. Select
VOLUME
using the
Keypad
.
The distribution volume is now
to be set to a correct value for
the individual patient. If desired,
it is possible for the machine to
carry out a distribution volume
calculation using the Watson
formula, see "Distribution
If the value is to be set manually,
when
VOLUME
has been
selected, set a proper value using
the
Display Up key
. 40 l is the
default starting point value .
Adjust the value as necessary.
Note that the distribution volume
has to be properly estimated
for the individual patient. The
calculation of Kt/V done in the
machine is entirely dependent
on the value for the individual
patient being the correct
one. For further information;
see "Glossary of Diascan®
parameters and key terms", page
The single clearance measurement check will be
performed on an appropriate occasion chosen by the
machine. The occasion chosen is dependent on when
the self-calibrations of the machine are to be performed.
DIASCAN
will be shown at the activity
fi
eld of the
Treatment Overview Menu when the function has been
activated.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Diascan
12:25
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...