UFR setting has been changed
Press SELECT key to confirm
Appears:
When treatment is resumed after the dialyzer and the
blood lines have been changed during treatment.
Machine actions:
Press the
Fluid Path button
, select
UF
and
PROFILING
using the
Keypad
. Check that the start
and stop values for
UF
are the correct ones, press the
Select button
to con
fi
rm this. If not, select
UF
and
deactivate pro
fi
ling. Change the values and reactivate
pro
fi
ling as usual procedure. See corresponding
instructions in chapter 7 "Pro
continues with set UF rate
Measures:
Mute time:
2 min.
Buzzer sound:
No buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:125
Attention alarms in alphabetical order