UF RATE STOPPED DUE TO HIGH PV
Lower PV
Appears:
When the set UF rate cannot be obtained due to too the
current high venous pressure.
Machine actions:
The dialysis
fl
uid is bypassed from the dialyzer.
Measures:
Check possible reason for the current high venous
pressure, e.g. the patient´s blood access, dialyzer clotting,
the functionality of the venous pressure transducer.
Correct accordingly. Make sure that the venous pressure
is decreased by at least 50 mmHg. The attention alarm
will automatically disappear when the current venous
pressure has been decreased.
If it is not possible to decrease the current venous
pressure, wait (maximum 5 minutes) until the dialysis
fl
uid enters the dialyzer again.
If the attention alarm reappears consider discontinuing
the treatment.
Mute time:
5 min.
Buzzer sound:
Soft buzzer
14:126
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order