Before the correct conductivity level has been reached the
fl
uid
fl
ow path will be orange. When the
fl
uid preparation is
fi
nished
the bypass path on the Flow Diagram lights up green.
7. Press the
fl
ashing
Fluid bypass button
.
The dialysis
fl
uid will now enter the dialyzer when the dialyzer
path on the Flow Diagram lights up green. Especially note that
the dialyzer should be
fi
lled from bottom to top to ensure proper
de-airation.
8. If the attention alarm
HIGH VENOUS PRESS TEST NOT MADE
appears, the
venous pressure has not exceeded 50 mmHg during the
priming procedure, which means that the machine has not
been able to test the venous pressure measurement function.
Establish a stable venous pressure over 50 mmHg until the
attention alarm disappears. If this attention alarm is left
unattended, unnecessary alarms and technical error will be
generated during initiating of treatment. See also “Venous
Patient Connector” in “Venous Blood Line - Attach”
previously in this chapter.
When the priming volume
has been achieved
,
the blood pump automatically stops and the
PRIMING VOLUME ACHIEVED
menu is displayed.
Check that the venous blood line is correctly placed in the
priming detector. See "Venous Blood Line - Attach", point 5.
It is now possible to continue priming (
EXTRA PRIMING
)
or if requested perform a recirculation procedure
(
RECIRCULATION
). It is also possible to go directly to patient
connection (
CONNECT PATIENT
).
Select desired alternative and continue to the corresponding
instruction as follows.
If necessary, select a new priming procedure (
NEW PRIMING
)
where priming will be repeated from the start and the set priming
volume value be set to zero.
4:42
AK 96
®
Operator's Manual - Double Needle Treatment
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...