Start the Treatment
CAUTION
Make sure that the blood path of the Flow Diagram lights up, if it
does not light up, check that the venous blood line is correctly placed
in the priming detector. If the blood path does not light up, the
machine considers the patient not connected and as a consequence
the UF system is not controlling the patient ultrafiltration.
If the air detector has not been activated during priming it will not be
automatically activated until the priming detector detects blood.
CAUTION
1. Check/adjust the blood
fl
ow rate to the required
value suitable for the particular patient using the
Blood Pump Up/Down keys
. Note especially that the blood
fl
ow rate in
SET QB
and
ACTUAL QB
, is the blood
fl
ow rate
when the blood pump is running in the arterial phase of the
single needle cycle. It is
not
the mean blood
fl
ow rate for the
single needle treatment. The mean blood
fl
ow rate is shown
on the Treatment Overview Menu for single needle treatment
"Overview Displays", see page 3:23 in part 1 for details.
2. If desired, press the
Blood Path button
and open the
BLOOD FLOW MENU
using the
Keypad
. Check/adjust the
low alarm limit for the blood
fl
ow rate, when the blood pump
is running in the arterial phase, in
SET LOW
as necessary.
3. Press the
fl
ashing
Arterial Pressure button
to centralize the
alarm limits around the current value. Or if desired, set the
arterial pressure alarm limits manually using the
Keypad
.
4. Press the
fl
ashing
Venous pressure button
. At that moment,
the high and low venous pressure alarm limits will
automatically be set. The high alarm limit will be set to +350
mmHg and the low alarm limit will be set to +150 mmhg.
5:8
AK 96
®
Operator's Manual - Single Needle Treatment
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...