Menus
The Diascan function have 2 menus displayed on the
Information Display; the Viewing Menu and the Setting
Menu.
Diascan measurement result values are to be read in the
viewing menu, which is displayed by
fi
rst pressing the
Fluid Path button
and then selecting
DIASCAN
using the
Keypad
. In this menu, select
SET DIASCAN
to open
the setting menu.
The Viewing Menu
When Diascan measurements have been performed,
the measurement result values for K and accumulated
Kt are displayed in this menu. A Kt forecast value is
also displayed. If Kt/V measurement is desired, and a
distribution volume has been set, the accumulated Kt and
the Kt forecast value will be replaced by Kt/V values.
Furthermore, if a Kt/V target value has been set, an
estimation of the time needed to reach the target is also
displayed in this menu. It is also possible to view the 20
last measurement results from this menu.
Measurement result values without Kt/V
measurement
-
LAST DIASCAN
shows
current time
for when the
Diascan measurement check was carried out.
12:8
AK 96
®
Operator's Manual - Diascan
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...