4.b.
Select
CITRIC 20%
.
5. Start the program by pressing the
Select key
for 3 seconds,
the
Rinse/Disinfection button
lights up. Note that if the
automatic start of heat disinfection or rinse program has been
activated, the
Rinse/Disinfection button
will already be lit.
DISINFECTION IN PROGRESS:
will be displayed and will
remain throughout the program.
The
Rinse/Disinfection button
will continue to be lit throughout
the program, indicating that the program is ongoing.
6. Connect the yellow disinfectant connector (see point 6 in
"Fluid Part Component Details", starting on page 2:19 in
part 1) to the pick-up tube and place it in the liquid citric
acid container, alternatively move the disinfectant connector
to the disinfectant port.
7. If the disinfectant connector is still connected to the parking
port when the
fi
lling phase of the program starts, the
following attention alarm will be generated:
Note especially that the disinfectant connector must remain in
the disinfectant container/disinfectant port and not be moved
to the parking port before the chemical disinfectant program is
complete. This is due that the chemical disinfectant is pushed
back to the disinfectant container to empty the connector/intake
tube at the end of the program.
8. If an automatic switch off is required, press the
On/Off button
for 3 seconds.
When the program is complete, drain sequence included, the
Priming button
starts to
fl
ash (not if an automatic switch off has
been done).
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Hygiene and Maintenance
8:17
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...