Integrated Heat Disinfection Program with WRO 300 H
- Performing
By an authorized technician, the AK 96 dialysis machine can be preset
to perform Integrated Heat disinfection program supported by the
WRO 300 H water puri
fi
cation unit. This program is recommended
when possible, since it includes heating of the inlet water tube.
The program starts with a heat disinfection program (or if selected, a
heat disinfection program in combination with CleanCart cartridge
or liquid citric acid) of the machine. When the circulation phase
(for information, see "Heat Disinfection - General" on page 8:9) has
fi
nished a low
fl
ow heat phase starts. In this phase, the machine will
receive hot water from the WRO 300 H, at a low
fl
ow rate, for 15
minutes . When this phase is complete, the machine
fi
nishes the
program off as usual, with a drain phase. The low
fl
ow heat phase
can be preset to be started automatically or manually, see following
handling instructions in this section.
Note
•
The program can be interrupted at any time when the program
is running; press and hold the
Select key
for 3 seconds. The
machine will then interrupt the procedure and
fi
nish off with
a drain sequence.
•
The pick-up tubes must be rinsed/disinfected separately. See
"Pick-up Tubes” on page 8:41 for instructions.
Note
Check before starting the procedure
That the concentrate connectors are placed in the corresponding
stand-by ports of the machine (blue and red), and that the yellow
disinfectant connector is placed in the parking port (marked P).
Make sure that the WRO 300 H is switched on.
That the dialysis
fl
uid tubes are connected to the safety couplings
and that the latches of the BiCart holder are closed.
Procedure
Use the
Keypad
next to the Information Display to select options and
to navigate within and between the menus. For details concerning the
keys included in the
Key pad
, see "Keypad" in chapter 3. part 1
The low
fl
ow heat phase can be started automatically or manually.
Automatic start is the default procedure described below, followed by
the manual start procedure.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Hygiene and Maintenance
8:19
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...