During the period of time when the measurement check
is being performed,
DIASCAN
, shown at the activity
fi
eld
of the Treatment Overview Menu, will be
fl
ashing.
When the measurement check has been performed, an
attention alarm text without a buzzer will appear on the
Information Display . Press the
Select key
to enter
the
LOG
menu. The measurement result values are also
displayed in the Diascan viewing menu.
When the measurement check has been performed
Diascan will automatically be turned to
OFF
. It is possible
to repeat the procedure at any time during treatment.
Note
•
To ensure a correct result a single clearance
measurement check should be activated no later than
30 minutes before discontinuing treatment. This is to
have enough time to obtain a measurement result.
Note
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Diascan