21.1. Clasp on the arterial blood line just before the pump
segment. Pull it out while simultaneously turning the
blood pump rotor in an anticlockwise direction. Make
sure that the pins on the blood pump rotor are behind
the pulled out segment while turning the rotor.
21.2. Open the cover of the air detector (press on the middle
of the cover at the same time as the cover is being
opened). Compress the venous drip chamber and then
pull it out as shown in the corresponding
fi
gures.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Double Needle Treatment
4:65