UF RATE LOWER THAN MINIMUM SET
Adjust TIME, UFV or MIN UF
Appears:
When the machine needs to decrease the UF rate below
the set minimum UF rate in order to obtain the set UF
volume at the end of treatment.
Machine actions:
Buzzer is activated 2 min. after attention alarm is
generated.
Measures:
Decrease the set minimum UF rate or increase the UF
rate by decreasing the treatment time, or by increasing
the UF volume.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
14:124
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order