DISCONNECT ARTERIAL PRESSURE
Press ARTERIAL button to confirm
Appears:
When the arterial pressure transducer of the arterial blood
line has been attached too early in function check. See
"Arterial Blood Line - Attach" on page 4:16 in part 1 for
information.
Machine actions:
None.
Measures:
Disconnect the arterial pressure transducer and press
Arterial button
to con
fi
rm.
If the attention alarm is still generated, consult an
authorized technician.
Mute time:
5 min.
Buzzer sound:
Soft buzzer
14:46
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order