HEPARIN PUMP OVERLOAD
Check cause
Appears:
When the pressure in the syringe is too high.
Machine actions:
The heparin pump stops.
Measures:
Check the syringe and the thin line connected to the
syringe for kinks. See Heparin Pump in "Arterial
Blood Line - Attach" starting on page 4:16 in part 1 for
instructions.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:69
Attention alarms in alphabetical order