Recovery from Machine Shut Down
In case of a short power failure, or a longer one where the machine
shuts down, the settings and accumulated values are stored by the
machine. If the machine has passed function check and time has
appeared on the Time Display before power failure, the machine will
perform a recovery when power returns. If the function check was not
completed before the power failure occurred, a function check will be
done when power returns.
When the machine performs a recovery an attention alarm will appear
and the buzzer sounds. The text “reC” is shown on the Time Display.
The operator must con
fi
rm the recovery after restart by pressing the
Select key
. The machine will then continue the treatment from where
it was interrupted.
CAUTION
Check treatment parameters after a recovery.
CAUTION
Note
•
If the attention:
appears, there will be no support for the buzzer and no
fl
ashing
segment on the Time Display during power failure. However,
the settings and accumulated values are stored by the machine.
When it has been con
fi
rmed, the machine will not restart
automatically.
The
On/Off button
has to be pressed to recover function.
•
The machine will still be able to operate since this is not the
same battery as the back-up battery. Contact an authorized
technician to change or recharge the battery.
•
During power failures, when the buzzer is muted, the current
consumption from the battery is high. In order to avoid
this high consumption, switch the machine off by pressing
the
On/Off button
for 3 seconds instead of pressing the
Mute button
. When the power returns, switch on again using
the
On/Off button
. A normal recovery will now follow.
Note
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Operating the Machine
3:29
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...