Conductivity Alarm
Appears:
The dialysis
fl
uid conductivity is outside the set alarm
limits.
Machine actions:
-
The dialysis
fl
uid is bypassed from the dialyzer.
-
If the dialysis
fl
uid conductivity is below 9 mS/cm,
the UF rate will automatically be set to zero.
-
When the dialysis
fl
uid conductivity is within the set
alarm limits again, the dialysis
fl
uid automatically
enters the dialyzer.
Possible cause:
-
The concentrate container/BiCart cartridge is empty.
-
Air is drawn into the machine instead of concentrate.
-
Incorrect concentrate has been connected e.g. the
connected concentrate is not in accordance with the
selected concentrate alternative.
-
There is an interruption in the concentrate
administration to the machine e.g. the concentrate
connector is not properly connected to the pick-up
tube, the BiCart cartridge is not properly attached
(see "Acidic concentrate and BiCart® cartridge" on
page 4:9 in part 1), the concentrate tube is kinked
or occluded, the
fi
lter integrated on the concentrate
tubes are clogged etc.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Alarms
13:45
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...