The Setting Menu
When
DIASCAN
has been selected in the viewing menu,
the setting menu will be displayed on the Information
Display. In this menu the Diascan function is to be
activated, measurement mode is to be chosen and, if
desired, an alarm limit for clearance can be set. If Kt/V
measurement is desired, distribution volume is to be set
in this menu. When this is done, it is possible to set a
target value and an alarm limit for Kt/V; these settings
are also entered here.
Activation/deactivation and measurement mode
setting
Choose a single measurement check or continuous checks
at intervals
(the intervals are 15, 30 or 60 minutes).
Activate the Diascan function by selecting
SINGLE
,
15 MIN
,
30 MIN
or
60 MIN
. Deactivate the function
by selecting
OFF
.
Note
If UFD is installed, the duration of the Diascan
measurement check is extended. When 15 minute
intervals have been set, the intervals may be
automatically extended to around 30 minutes.
Note
12:12
AK 96
®
Operator's Manual - Diascan
HCEN12239 Revision 12.2010
Program version 3.xx