TOO HIGH TMP REQUIRED
Increase time or decrease UFV
Appears:
When the machine cannot maintain the required UF rate.
Machine actions:
Buzzer is activated 2 min. after attention alarm is
generated.
Measures:
Check the dialyzer; too small/partly clotted, and do
necessary actions in accordance with this. If possible,
adjust set Time and/or set UF volume until TMP
decreases.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:119
Attention alarms in alphabetical order