4.
Parking Port marked P, for Yellow Disinfectant
Connector
The yellow disinfectant connector must be connected to this
yellow port (marked P) at all times except when a chemical
disinfection program is performed. See more information in
point 6 further on in this list.
5.
Disinfection Port for Yellow Disinfectant Connector
The yellow disinfectant connector (point 6 in this list)
is to be connected to this yellow port when a chemical
disinfection program is performed. This is on condition that
the disinfectant used for chemical disinfection program is
permanently connected to the inlet line at the back of the
machine (see "The Rear of the Machine" on page 2:28 in
the next section).
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Machine Description