BLOOD PATH PRESSURE TEST FAILED
Press SELECT key to continue
Appears:
May occur during assisted priming if the automatic
pressure test/leakage test of the dialyzer fails.
Machine actions:
The blood pump stops.
Measures:
Press
Select key
. The priming phase will start
automatically when the blood pump is started.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:19
Attention alarms in alphabetical order