Measures:
-
Press the
fl
ashing
Alarm button
and the alarm text
will be shown on the Information Display.
-
Check that the venous blood line is properly inserted
into the priming detector.
When the priming detector detects blood in the
venous blood line, the blood path of the Flow
Diagram lights up red and the treatment time starts to
count down on the Time Display.
-
Press the
Select key
to con
fi
rm.
-
When done select
CONTINUE CONNECT PATIENT
and press the
Select key
.
13:50
AK 96
®
Operator's Manual - Alarms
HCEN12239 Revision 12.2010
Program version 3.xx