LOW FLOW HEAT NOT AVAIL. IN WRO
Press SELECT key to confirm
Appears:
When attempting to start integrated heat disinfection
program with WRO 300 H, without a WRO model which
supports low
fl
ow heat (where the machine receives hot
water from the WRO 300 H).
Machine actions:
None.
Measures:
Press the
Select key
. to con
fi
rm.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Supersoft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:91
Attention alarms in alphabetical order