Dialysis Fluid Tubes - Attach
The dialysis
fl
uid tubes may be attached to the dialyzer when the
function check is complete, i.e. "F.Ch" disappears from the Time
Display and the bypass path of the Flow Diagram lights up. The
Fluid bypass button
will start to
fl
ash and when pressed, the supply
of dialysis
fl
uid will be activated.
The dialysis
fl
uid preparation is complete (correct conductivity level
has been reached) when the bypass path on the Flow Diagram lights up
green. If the dialysis
fl
uid tubes are attached to the safety couplings of
the machine at that moment, the dialysis
fl
uid will start to circulate in
the dialysis
fl
uid tubes. If the dialysis
fl
uid tubes have been attached to
the dialyzer, the dialysis
fl
uid will enter the dialyzer.
Note
•
If the optional Dialysis Fluid Filter (UFD) has been installed,
the dialysis
fl
uid tubes must remain connected to the safety
couplings until the bypass path of the Flow Diagram lights up
green. The machine performs a
fl
ush of the
fl
ow path where the
ultra
fi
lter is placed after “F.Ch” has disappeared from the Time
Display. This is to ensure that correct conductivity is reached
when the dialysis
fl
uid enters the dialyzer. There is a large
volume of
fl
uid passing through the dialysis
fl
uid tubes during
the
fl
ush. An attention alarm text is simultaneously shown on
the Information Display. Wait until the attention alarm text has
disappeared and the bypass path of the Flow Diagram lights up
green before disconnecting the dialysis
fl
uid tubes
1
.
Note
1
If assisted priming is performed disregard note this note
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Double Needle Treatment
4:35
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...