Residual Test after Chemical Disinfection
Before connecting to a patient, a residual test must be done on the
prepared dialysis
fl
uid. This is to make sure that the patient will not be
exposed to chemical residuals after a chemical disinfection program.
The residual concentration in the dialysis
fl
uid must be below levels
speci
fi
ed by the clinic or by national standards. It is essential to use an
appropriate test method, either with proven sensitivity for the chemical
or recommended by the manufacturer of the chemical disinfectant.
CAUTION
After the Chemical disinfection procedure a test for residues must
have been performed prior to connecting to a patient.
CAUTION
Note
•
The pick-up tubes must be rinsed/disinfected separately. See
"Pick-up Tubes” on page 8:41 for instructions.
Note
Procedure
1. Switch the machine on and let it pass function check with the
proper concentrates attached.
2. When the green
fl
uid path lights up, press the
fl
ashing
Fluid bypass button
in order to let fresh dialysis
fl
uid
fl
ow
through the dialysis
fl
uid tubes.
3. Take the residual test sample from the
fl
owing
fl
uid on the
dialysis
fl
uid outlet tube (to dialyzer).
If the test shows disinfectant residues:
•
DO NOT use connected BiCart cartridge.
•
Press the
Rinse/Disinfection button
and start a new rinse/drain
procedure (the button lights up). When this is
fi
nished, repeat
the handling procedure described above.
If the machine still shows disinfectant residues, contact an authorized
technician.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Hygiene and Maintenance
8:31
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...