Ultrafilter - How to change
Note
•
Make sure that the ultra
fi
lter is changed in an aseptic way.
Note
Procedure
1. Release the lower part of the ultra
fi
lter holder by pulling the handle
and pressing the latch downwards.
2. Remove the ultra
fi
lter by pulling it gently downwards.
3. Insert the new ultra
fi
lter into the holder and push it gently upwards.
4. Close the holder by pushing the lower latch into position.
5. Con
fi
rm the change of ultra
fi
lter (see Ultra
6. Label the ultra
fi
lter with date for change.
7. Perform a Heat disinfection program before the AK 96 dialysis
machine is used for treatment after the ultra
fi
lter has been changed.
Ultrafilter replacement reminder
If UFD (Ultra
fi
lter Dialysis Fluid) is installed the operator will
be reminded to change ultra
fi
lter. The operator will be noti
fi
ed
by an attention alarm when a limit has been exceeded. There are
three different limits: number of days since last change, number
of heat disinfections since last change and number of hypochlorite
disinfections since last change.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Hygiene and Maintenance
8:39
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...