4.a.
Select
CHANGE
.
4.b.
Select
HEAT
.
5. Start the program by pressing the
Select key
for 3 seconds,
the
Rinse/Disinfection button
lights up. Note that if the
automatic start of heat disinfection or rinse program has been
activated, the
Rinse/Disinfection button
will already be lit.
DISINFECTION IN PROGRESS:
will be displayed and will
remain throughout the program.
The
Rinse/Disinfection button
will continue to be lit throughout
the program, indicating that the program is ongoing.
6. If an automatic switch off is required, press the
On/Off button
for 3 seconds.
When the program is complete, drain sequence included, the
Priming button
starts to
fl
ash (not if an automatic switch off has
been done).
7. Switch the machine off by pressing the
On/Off button
for
3 seconds.
If a new treatment is to be started directly after the disinfection
program, press the
fl
ashing
Priming button
brie
fl
y and
PRESS SELECT 3 SECONDS TO START FCH
will be
displayed. Press the
Select key
for 3 seconds to start a new
function check and simultaneously reset previous values to zero.
8:12
AK 96
®
Operator's Manual - Hygiene and Maintenance
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...