
BLOOD LEAK DETECTOR FAILURE
Clean and press SELECT key
Appears:
During function check if the blood leak detector is dirty
or out of function.
Machine actions:
None.
Measures:
Clean the blood leak detector, see "Blood Leak Detector"
on page 8:41 in part 1 for instructions. Press the
Select key
when cleaned.
If the attention alarm appears again, call an authorized
technician.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:17
Attention alarms in alphabetical order
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...