Disinfection History
The disinfection history list can not be shown during treatment.
At all other times when the machine does not detect blood, the
Rinse/Disinfection button
can be pressed for 3 seconds and a list of
the 3 most recently performed disinfection programs (both heat and
chemical disinfection programs) will be shown on the Information
Display.
The last performed disinfection program is shown at the top of the
list. The 20 most recently performed disinfection programs can be
displayed by scrolling up and down using the
Display Up key
and the
Display Down key
.
If the disinfection program was
successfully
performed;
COMPL
is
shown to the right on the Information Display. The time shown to the
left is the time for when the disinfection program was
completed
.
If the disinfection program was
not successfully
performed;
INCOMP
is
shown to the right on the Information Display. The time shown to the
left is the time for when the disinfection program was
started
.
Go back from the disinfection history list by pressing the
Back key
.
8:4
AK 96
®
Operator's Manual - Hygiene and Maintenance
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...