
VENOUS LOW LIMIT BELOW +10 mmHg
Press VENOUS button to adjust
Appears:
10
minutes after the venous low alarm limit has been
set below + 10 mmHg.
Machine actions:
None.
Measures:
Press
Venous Pressure button
and adjust the alarm limit
to a proper value.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Continuous buzzer
14:130
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...