Blood Pressure Monitor (BPM)
Blood pressure monitoring is only available if the optional BPM is
installed.
The alarm limits below can be preset. The value put in brackets and in
italics is the default value.
Systolic pressure range
60 - 250 mmHg
Low alarm limit
60 - 250 mmHg
(100 mmHg)
High alarm limit
60 - 250 mmHg
(150 mmHg)
Diastolic pressure range
40 - 200 mmHg
Low alarm limit
40 - 200 mmHg
(60 mmHg)
High alarm limit
40 - 200 mmHg
(100 mmHg)
Mean pressure range
45 - 235 mmHg
Low alarm limit
45 - 235 mmHg
(75 mmHg)
High alarm limit
45 - 235 mmHg
(115 mmHg)
Pulse rate range
40 - 200 bpm (±3 bpm or ±2 % of
reading)
Low alarm limit
40 - 200 bpm
(40 bpm)
High alarm limit
40 - 200 bpm
(150 bpm)
1
Meets ANSI/AAMI SP-10 (1992). Mean error ±5 mmHg. Standard deviation
8 mmHg
9:4
AK 96
®
Operator's Manual - Technical Data
HCEN12239 Revision 12.2010
Program version 3.xx