Blood Path button
Press the
Blood Path button
to reach the Blood Path main menu. In
this menu, the
BLOOD FLOW
can be read/set/adjusted. Settings and
activation of the
HEPARIN
pump are to be made and settings and
activation for the
SINGLE NEEDLE
function can be made.
The following will be shown on the Information Display:
BLOOD FLOW
; current blood
fl
ow rate
HEPARIN
; sum of accumulated bolus volume during priming and total
accumulated volume since treatment start
SINGLE NEEDLE
; current stroke volume
Blood Flow
From the Blood Path main menu; select
BLOOD FLOW
by using the
Keypad
to reach the sub menu.
The following will be shown on the Information Display:
SET QB
; set blood
fl
ow rate
SET LOW
; set low alarm limit for blood
fl
ow rate
SEGMENT
; blood pump segment diameter
ACTUAL QB
; current blood
fl
ow rate
ACC
; accumulated blood volume since treatment start
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Operating the Machine
3:13