Command Reference
R&S
®
ZNL/ZNLE
871
User Manual 1178.5966.02 ─ 07
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number. Segment numbers must be sequential.
The specified segment number must be smaller or equal to the
number of existing segments plus 1.
If segment number
<Seg>
already exists, it is replaced by the
new segment.
Parameters:
<StartFreq>
Start frequency of the segment; see
.
Default unit: Hz
<StopFreq>
Stop frequency of the segment; see
.
Default unit: Hz
<Points>
Number of sweep points in the segment. See
.
<Power>
Internal source power in the segment. See
.
Default unit: dBm
<SegmentTime>
Duration of the sweep in the segment. See
DWELl
, this parameter is
replaced by
<MeasDelay>
.
Range:
Depending on other channel settings. AUTO acti-
vates automatic sweep time setting in the segment,
which is equivalent to the minimum sweep time
possible.
Default unit: s
<MeasDelay>
Delay for each partial measurement in the segment. See
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl
. In the setting
[SENSe<Ch>:]SEGMent<Seg>:INSert:SELect
SWTime
,
this parameter is replaced by
<SegmentTime>
.
<Unused>
Ignored parameter (for compatibility with R&S ZVR analyzers).
Should be set to the default value 0.
<MeasBandwidth>
IF bandwidth in the segment. See
]SEGMent<Seg>:BWIDth[:RESolution]
.
Default unit: Hz
<LO>
POSitive | NEGative | AUTO
Position of the local oscillator frequency LO relative to the RF
frequency. In remote control this parameter must be set when a
sweep segment is created. See
.
VNA Command Reference