VNA GUI Reference
R&S
®
ZNL/ZNLE
446
User Manual 1178.5966.02 ─ 07
Detect Port Assignment
Starts a procedure by which the R&S
ZNL/ZNLE (with a little help from the attached
calibration unit) auto-detects the connected ports. The automatic assignment replaces
the configured one.
In case auto-detection fails
●
an error report is shown as a warning dialog
●
the undetected port connections are marked with warning signs
●
the calibration can be invalid
Remote command:
[SENSe:]CORRection:COLLect:AUTO:PORTs:CONNection?
Apply/Cancel
Apply the calculated system error correction to the active channel (or to all channels in
the active channel setup, if all channels were calibrated).
The Apply button is enabled when calibration sweeps have been successfully per-
formed for all required port assignments.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
Extended cal unit settings:
MMEMory:AKAL:FACTory:CONVersion
MMEMory:AKAL:USER:CONVersion
SYSTem:COMMunicate:AKAL:CONNection
SYSTem:COMMunicate:AKAL:MMEMory[:STATe]
9.12.1.4
Calibration Setting Wizard
The "Calibration Setting" wizard guides you through the setup and execution of a man-
ual system error correction.
Access:
[Cal] > "Calibration" > Calibration – "Start... (Manual)"
The wizard proceeds through the following steps:
1.
"Ports and Type":
Select the ports you want to calibrate and the calibration type
you want to use.