Command Reference
R&S
®
ZNL/ZNLE
920
User Manual 1178.5966.02 ─ 07
Parameters:
<SearchValue>
Target search value of marker no.
<Mk>
.
Range:
Depending on the format of the active trace. For a
dB Mag trace the range is -200 dB to +200 dB.
Increment: 0.1 dB
*RST:
Depending on the trace format; 0 dB for a dB Mag
trace.
Default unit: dB
CALCulate<Chn>:MARKer<Mk>:FUNCtion[:SELect]
<Mode>
Selects a search mode for marker no.
<Mk>
, which can then be initiated using one of
the
CALCulate<Ch>:MARKer<Mk>:SEARch...
,
CALCulate<Ch>:MARKer<Mk>:MAXimum
or
CALCulate<Ch>:MARKer<Mk>:MINimum
functions. The marker must be created
before using
CALCulate<Chn>:MARKer<Mk>[:STATe]
.
Note:
This command is not needed except for compatibility with ZVR programs. Use
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
to select a search mode and
at the same time initiate the search. The
CALCulate<Ch>:MARKer<Mk>:SEARch...
,
CALCulate<Ch>:MARKer<Mk>:MAXimum
or
CALCulate<Ch>:MARKer<Mk>:MINimum
functions also select the search mode.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk>
Marker number. For a bandfilter search (
BFILter
) this numeric
suffix is ignored and may be set to any value because the band-
filter search functions always use markers M1 to M4.
Parameters:
<Mode>
MAXimum | MINimum | RPEak | LPEak | NPEak | TARGet |
LTARget | RTARget | BFILter | MMAXimum | MMINimum |
SPRogress
See
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
CALCulate<Chn>:MARKer<Mk>:MAXimum
CALCulate<Chn>:MARKer<Mk>:MINimum
These commands select a search mode for marker no.
<Mk>
and initiate a maximum
and minimum search, respectively. The marker must be created before using
CALCulate<Chn>:MARKer<Mk>[:STATe]
ON
.
Note:
This commands are the ZVR-compatible equivalent of
MINimum | MAXimum
.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk>
Marker number.
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH