Command Reference
R&S
®
ZNL/ZNLE
912
User Manual 1178.5966.02 ─ 07
Example:
See
TRIGger<Ch>[:SEQuence]:MULTiple:SOURce
Manual operation:
See
TRIGger<Ch>[:SEQuence]:MULTiple:SOURce
<Sequence>[, <TrigSource>]
Selects the source of the trigger events the analyzer uses to start a measurement
sequence in multiple trigger mode (
TRIGger<Ch>[:SEQuence]:SOURce MULTi
).
Suffix:
<Ch>
.
Channel number
Parameters:
<Sequence>
SWEep | SEGMent | POINt | PPOint
Triggered measurement sequence
PPOint
denotes "partial measurement"; see
.
<TrigSource>
IMMediate | EXT1 | EXT2 | E1A2 | E1O2 | MANual
Trigger source for the triggered measurement sequence
IMMediate
Free run measurement (untriggered)
EXT1
Trigger event at Trigger In or Pin 2 of
The latter requires option R&S
FPL1-B5 (R&S
ZNL only).
EXT2
.
Requires option R&S
FPL1-B5 (R&S
ZNL only).
E1A2
Trigger events at EXT1
AND
EXT2
E1O2
Trigger event at EXT1
OR
EXT2
MANual
Trigger event generated by pressing the "Manual Trigger" soft-
key
Example:
TRIG:MULT:SOUR SWE, EXT1; SOUR POIN, EXT2
Select external trigger 1 as a trigger source for the entire sweep,
external trigger 2 as a trigger source for each point.
TRIG:MULT:SLOP1 SWE, POS; SLOP2 POIN, POS
Trigger on the rising edges of the external trigger signals 1 and
2.
TRIG:MULT:HOLD POIN, 1ms
Define a trigger delay of 1 ms before each sweep point.
Manual operation:
See
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH