VNA GUI Reference
R&S
®
ZNL/ZNLE
502
User Manual 1178.5966.02 ─ 07
If the "Fixture Simulator" is disabled for the related channel (see
on page 471), this tab is inactive, i.e. all controls except the "Overview" button are
grayed out.
Overview
See
Type
Currently only Differential Match"Embedding" is supported.
Logical Port
Logical analyzer port, as defined in the "Balanced Ports" configuration. The transfor-
mation networks are defined such that the physical analyzer test ports are connected
to the left of the circuit; the DUT ports are on the right side. You can define independ-
ent embedding networks for all balanced ports.
Remote command:
The
<LogPt>
numeric suffix in the embedding/deembedding commands identifies the
CALCulate<Ch>:TRANsform:VNETworks:DIFFerential:
Network
The graphical list contains the available 2-port networks for Differential Match embed-
ding:
●
The symbol selects "no network" and disables differential match embedding for
the selected balanced port.
●
The "2-Port Data" network is defined by means of imported S-parameter data; see
●
The "Shunt L, Shunt C" network is defined by lumped elements whose parameters
are displayed below the graphical list.
Offset Embed Softtool
www.allice.de
Allice Messtechnik GmbH